PRODUCT SPECIFICATIONS
PERFORMANCE
Flow Range ..................... 1- 5000 cc/min (Total),
850 - 5000 cc/min (High Flow);
5-500 cc/min (Constant Low Flow);
1-750 cc/min (Constant Pressure)
Constant Flow Control ..... Better than ± 5% of set flow
(after calibration)
Pressure Range .............. 5000cc up to 8” H20 back pressure
4000cc up to 15” H20 back pressure
3000cc up to 23” H20 back pressure
2000cc up to 26” H20 back pressure
1000cc up to 25” H20 back pressure
850cc up to 20” H20 back pressure
Run Time ........................ 8 hour minimum
Flow Fault ....................... If flow changes exceed 5%, fault
icon appears. If fault exceeds 30
seconds, pump shuts down. Pump
attempts to restart every 3 minutes
for up to 1 hour
ENVIRONMENTAL
Temperature Ranges
Operating ..................... 32°F to 104°F (0°C to 40°C)
Storage ........................ -49°F to 113°F (-45°C to 45°C)
Charging ...................... 41°F to 104°F (5°C to 40°C)
Humidity Ranges
Operating ..................... 0-85 %RH, non-condensing
Storage ........................ 0-100 %RH, non-condensing
GENERAL
Display Messages ........... LAST, CAL, SHUT/OFF (Clock, Program);
E, PC, P1-P6, OLD (Program)
Controls .......................... Power Switch, Flow Control Screw,
s & t Buttons (Clock, Program),
Programming (PROG) Button (Program)
Indicators ........................ Elapsed Time (Clock, Program)
Icons (LCD) ..................... Low Battery, Flow Fault, Clock
(Clock & Program)
Dimensions ..................... 4.1W x 3.9H x 2.0D inches
10.4W x 9.9H x 5.1D cm
Weight ............................ 22.5 oz. (638 g)
ELECTRICAL
Battery Pack ................... Removable, Sealed,
Rechargeable Nickel-Cadmium
Interface Connectors ....... Charging Jack
Charging Time ................ 14-18 hours
APPROVALS
CE .................................. EMC EMI/RFI, EN 55 022 Class B;
IEC 801-2, 3
Intrinsic Safety
UL .................................. Class I, Div 1, Groups A, B, C, D;
T3C
CENELEC ........................ EEx ia IIC T4
• External Filter Housing
Clear External filter housing allows an
immediate visual inspection of filter’s
condition.
• Flow Fault & Test Indicators
Flow Fault indicator lights up if unit
operates outside ± 5% of set flow rate.
Battery Test indicator lights up when
battery can provide a minimum of 8
hours of operation over operating
range.
• Basic, Clock, & Program Models
Basic model is easy to use. Clock
model provides elapsed-time clock
display. Program model provides delay
start times up to 9999 mins.
• Optional Low Flow Modules
Constant Low Flow module allows
sampling from 5-500 cc/min. Constant
Pressure (Multiflow) Module allows
sampling from 1-750 cc/min.